Air Compressor System
The medical compressed air plant provides a reliable and constant air supply, without the transport and management costs associated with heavy gas cylinders. A wide range of models are available to satisfy all sites of installations. The plants are designed to operate at a maximum of 10 bar. All range includes duplex components to ensure maximum plant availability. Air delivery filtered by pre-filter, oil coalescing filter, activated carbon filter, and a dust filter to give medical air of a quality as defined ISO 8573.1. An additional sterile filter (doubled) is provided. The medical Air central station incorporates the following elements: Three identical air electro compressors, screw type Air receivers according with compressors load and protected by a pressure relief valve Filter combination. Air delivery is via twin pre-filters, activated carbon filters, dust filters and sterile filters. Absorption or refrigerator double twin column dryer assembly operated by its own control panel. Control system contains all elements necessary for pressure control, duty selection, indication and alarm functions and pressure head out. Distance alarm.